X-ray devices, System X-ray diagnostic "OPERA" on three workplaces with system of digital processing of the image
- BrandGeneral Medical Merate S.p.A.,
- Country of manufactureItaly
The x-ray device on three workplaces on the basis of a long-distance table support.
Scopes: gastroenterology, miyelografiya, pediatrics, urology and gynecology, traumatology, emergency medicine.
Operational modes: digital X-ray analysis, tomography, roentgenoscopy.
Technical characteristics
Name of parameter |
Value |
Remotely-controlled table |
|
Table inclination range |
+90 °/-30 ° |
The maximum speed of the movement at an inclination |
6 ° / sec. |
Automatic stop of a table in horizontal position |
existence |
Range of cross movement of a sound board of a table with automatic centering |
34 cm |
Range of longitudinal movement of a column and detector/URI |
124 cm with a changeable speed up to 20 cm / with |
The focal length with continuous fine tuning changed in limits |
115 ÷ 150 cm |
Distance "a table sound board – a floor" |
86 cm |
Management of all movements of a complex directly from the console located on a table |
existence |
Table sound board loading capacity |
182 kg (227 kg with restrictions) |
Free access from all directions |
existence |
Column inclination range with automatic centering on object of research |
± 40 ° |
Sizes: - table sound boards - rentgenprozrachny zone of a sound board of a table |
210х74 cm 194х50 cm |
Min. distance from the center of radiation to the brim a table |
43 cm |
Min. distance "focus - the detector" |
78 mm |
Min. distance "patient URIAH" |
138 mm |
Turn of a tube |
270 ° |
The eliminating lattice |
existence |
Compression device |
|
Remotely-controlled, motorized, with the autoparking |
existence |
Adjustable effort of a compression |
from 50 to 200 N |
Collimator |
|
Iris, automatic fine tuning taking into account focal length, deduction function |
existence |
The size of the field of collimation at focus of 100 cm |
43х43 cm |
Tomography |
|
Bidirectional linear tomography |
existence |
It is carried out in any provision of a table |
existence |
Tomography corners |
8 °, 20 °, 40 ° |
Tomography speeds |
2 |
Height of a tomographic layer with electronic adjustment |
from 0 to 330 mm, with an adjustable step of 1 ÷ 10 mm |
Automatic increase in height of a layer |
existence |
Display of management of a table and generator |
integrated |
Amplifier of the x-ray image |
|
Nominal size of the entrance field |
9 inches |
Nominal diameter of fields |
9-6-4 inches |
CCD camera |
1024x1024x12 |
Automatic adjustment of strengthening of video signal |
existence |
Dynamic recursive filter |
existence |
Movement detector |
existence |
System of digital processing of the image with the flat panel detector |
|
Technology |
Flat panel, photodiode matrix |
Sensitive surface |
43kh43sm |
Matrix |
2880х2881 pixels |
Pixel size |
143 microns |
Processing of the image: control of brightness and contrast, inversion of a gray scale, spatial filters, collimation, ZOOM, screen magnifying glass, rotation of the image, texts. Measurements: distances, corners. |
existence |
Uniform control interface of the generator, table and system of digital processing of the image |
existence |
DICOM 3: STORE, PRINT, Worklist, DVD with record of the browser on each disk for the subsequent reproduction on any personal computer |
existence |
ZhK monochrome monitor 19" |
existence |
X-ray radiator |
|
X-ray tube with the rotating anode |
20/50 kW; 150 kV |
Amount of focuses |
0,6/1,2 mm |
Anode thermal capacity |
300 kHu |
The high-frequency generator with the control panel |
|
Power |
50 kW |
Frequency of a high voltage |
200 kHz |
Range of output tension: - X-ray analysis - roentgenoscopy |
40-150 kV with a step of 1 kV 40-125 kV with a step of 1 kV |
Current range - X-ray analysis - roentgenoscopy |
10-600 ma 0,5-10 ma |
Exposition time at a X-ray analysis |
0,001-6,3 sec. |
Exposition time at roentgenoscopy |
0-5-10 min. |
Range Mas |
0,1-1000 Mas |
Technology of shooting |
0,1,2,3 t. |
Automatic control of an exposition |
existence |
Automatically adjustable roentgenoscopy |
existence |
Protection: an overload on current, tension, control of existence of rotation of the anode and temperature of the anode |
existence |
Self-calibration on current |
at each exposition |
Exposure meter |
existence |
Accessories: set of high-voltage cables, hand-rail, footboard, emphasis for shoulders, a frame for uniform distribution of weight of the device, glass X-ray protective 100kh80sm 2,8Pb, the laser multiformat printer with a starting cartridge of a rentgenplenka. |
For simultaneous work the X-ray of the laboratory assistant and a X-ray of the doctor need to be established an additional workplace.